RTR-4 Operator’s Manual
System Description, Setup, and Quick Start
120300 Rev. D
SAIC Proprietary
3-19
Procedure RTR0004
X-Ray an object as follows:
STEP
ACTION
WARNING
Excessive exposure to X-ray radiation is harmful.
Verify the exclusion zone is clear of all personnel.
Failure to comply may result in personnel injury.
1
Press the
F1
key or click the
Acquire
button
on the toolbar. The
Acquire
dialog box appears.
2
Enter the number of pulses desired and the desired
time delay.
3
Press the
Enter
key or click on the
Fire X-Ray
button. The source fires and an image appears on the
screen.
NOTE
The X-ray source fires and an image is produced.
Examine image to determine if the image shall be
saved or discarded. Press
F2
to
QuickSave or F4
to
Save As
or if the image is too dark, discard it and
take another image, either with a higher pulse
settings or use the Sum feature. If the image is too
light, discard it and take another image with a lower
pulse setting. Proceed to the appropriate procedure in
the operations chapter.
RTR0004a
RTR23
RTR0004b