RTR-4 Operator’s Manual
System Description, Setup, and Quick Start
120300 Rev. D
SAIC Proprietary
3-7
The functions located on the Imager control panel are as follows:
CONTROL OR
INDICATOR
FUNCTION
Power On-Off Switch
Turns on the imager power for both
internal battery and external power
operation.
Power On LED
• Green and on steady when
powered by battery.
• Amber and on steady when
powered by an external source.
• Flashing green when power-up
delay is selected, with last five
seconds before power on rapid
flashing green.
• Flashing Amber when battery
power is low.
Battery Charger LED
• Amber and flashing when testing
battery.
• Amber and on steady during
battery charging.
• Green and on steady when battery
charging is complete.
Data Connector
Receptacle for the Extension Data
Cable Spool. Color coded green to
match the cable connector.
X-Ray Trigger
Connector
Receptacle for the X-Ray Trigger
Cable. Color coded yellow to match
cable connector.
External Power
Connector
Receptacle for an external 12Vdc to
24Vdc power source.
Battery
Compartment
Receptacle for the NiMH 10.8Vdc
battery. Insert battery with the “This
side up” label towards the external
power connector.
Power On Delay
Timer
Provides for setting power-on delays
from 0 to 3 minutes in 20 second
intervals.