51
Reference Manual
00809-0100-4803, Rev GA
Configuration
September 2017
Configuration
If conversion factor tables or internet search engines are used to determine the Base per Custom value, it
is important to enter the Custom Unit in the “From” field and the Base Unit in the “To” Field.
To calculate the Base per Custom value for a custom unit not shown in
example:
Totalizer conversion example
To find the Base per Custom relationship for a custom unit of millions of standard cubic feet (mmscf) and
a base unit of standard cubic feet (StdCuft), input the following:
Custom Unit = mmscf
Base Unit = StdCuft
Because:
1 mmscf (Millions of Standard Cubic Feet) = 1000000 StdCuft (Standard Cubic Feet)
Therefore:
Base per Custom =
=
= 1000000
Figure 2-35. Totalizer Custom Units - Totalizer Example
Number of Base Units
1
x
Custom
x
Unit
----------------------------------------------------------------------
1000000 StdCuft
⋅
1 mmscf
⋅
-------------------------------------------------