![Rosemount 3051S Series Reference Manual Download Page 49](http://html.mh-extra.com/html/rosemount/3051s-series/3051s-series_reference-manual_1487385049.webp)
41
Reference Manual
00809-0100-4803, Rev GA
Configuration
September 2017
Configuration
Mass/volume conversion example
To find the Base per Custom relationship for a custom unit of kilograms per hour (kg/h) and a base unit of
grams per hour (g/h), input the following:
Custom Unit = kg/h
Base Unit = g/h
Because:
1 kg (Kilogram) = 1000 g (Grams)
Then:
1 kg/h =
⫻
= 1000 g/h
1 kg/h = 1000 g/h
Therefore:
Base per Custom =
=
= 1000
Figure 2-28. Flow Rate Custom Units - Mass/Volume Conversion Example
1 kg
⋅
1 h
⋅
--------------
1000 g
⋅
1 kg
⋅
---------------------
Number of Base Units
1
x
Custom
x
Unit
----------------------------------------------------------------------
1000 g h
⁄
⋅
1 kg h
⁄
⋅
----------------------------
Summary of Contents for 3051S Series
Page 2: ......
Page 126: ...118 Reference Manual 00809 0100 4803 Rev GA Troubleshooting September 2017 Troubleshooting ...
Page 189: ......