Preparing for Use
Read the actual the Firmware Version
Quick Start Guide 1505.3896.62 - 14
31
2.12.2 Get Firmware Version via SCPI Command
In remote mode the current OSP Firmware Version can be read via the “*IDN?”
command. When sending the Identification command, the OSP prompts back the
following answer (example for OSP120) after pressing the Read button (see below
example).
The last part of the string shows the actual firmware version (1.5 for the above
example).
With Firmware version 2.51 onwards, the actual version of the OSP is read-out in the
configuration string, too.
See below example for OSP120 with Ser. No. 100008 and Var 02:
Rohde&Schwarz, OSP120, 100008, Var02, 2.51
With Firmware version 2.57 onwards, the firmware version of the OSP is read-out with
additional date label.
See below example for OSP120 with Ser. No. 100008, Var 02 and date 2016, June
14
th
:
Rohde&Schwarz, OSP120, 100008, V02, 2.57.160614