Instrument Tour
R&S
®
EVSF1000
32
Getting Started 1178.6404.02 ─ 05
5.2.9
AF Out (R&S
EVSF1-B4 Only)
Provides AF output to connected headphones or a loudspeaker (optimized for
8
Ω
). When using the R&S
EVSF1-Z2 service adapter, audio output is sent to the
speaker.
5.2.10
Trigger In (R&S
EVSF1-B4 Only)
Provides an external trigger for data recording. The voltage level is 3.3
V to 12
V.
The typical input impedance is 1
MΩ.
When using the R&S
EVSF1-Z2 service adapter, use the BNC connector, 50
Ω
/20k
Ω
.
5.2.11
IP-Address Select (R&S
EVSF1-B4 Only)
Determines which of two possible IP addresses is used, to distinguish between
two instruments in a single rack.
If this pin is not connected, the address configured as "TCP/IP Address 1" is
used.
If this pin is connected to ground, the address configured as "TCP/IP Address 2"
is used.
5.2.12
Device ID
The unique device identifier is provided as a barcode sticker on the rear panel of
the R&S
EVSF1000.
It consists of the device order number and a serial number.
Rear Panel