Preparing for Use
R&S
®
EVSF1000
23
Getting Started 1178.6404.02 ─ 05
To perform an auto-calibration, you require remote access to the instrument via
LAN. For details see
To perform an auto-calibration
Before performing an auto-calibration, make sure that the instrument has reached
its operating temperature (after about 15 minutes of operation; for details, refer to
the data sheet).
A message in the status bar ("Instrument warming up...") indicates that the oper-
ating temperature has not yet been reached.
During the auto-calibration procedure, do not apply any input signal to the
R&S
EVSF1000.
1. Using the VNC viewer, select [n] > [F8] > [F1]
2. Select "Start Autocal".
3. Press [Enter].
The "Setup - CAL" display shows the status for each calibration step. When
completed, after a few minutes, all entries must have the status "OK".
For more information on auto-calibration, see the R&S
EVSF1000 user manual.
Performing an Auto-Calibration