Preparing for Use
R&S
®
FSVA/FSV
43
Quick Start Guide 1321.3066.02 ─ 06
4.1.9
Performing a Self Alignment and a Self Test
Operating temperature
Before performing this functional test, make sure that the instrument has
reached its operating temperature (for details, refer to the data sheet).
Performing a self alignment
1. Press the SETUP key.
2. Press the "Alignment" softkey.
3. Press the "Self Alignment" softkey.
Once the system correction values have been calculated successfully, a mes-
sage is displayed.
To display the alignment results again later
●
Press the SETUP key.
●
Press the "Alignment" softkey.
●
Press the "Show Align Results" softkey.
Performing a self test
The self test does not need to be repeated every time the instrument is switched
on. It is only necessary when instrument malfunction is suspected.
1. Press the SETUP key.
2. Press the "More" softkey.
3. Press the "Service" softkey.
4. Press the "Selftest" softkey.
Once the instrument modules have been checked successfully, a message is
displayed.
Putting into Operation