Controlling the R&S
FSVA/FSV Remotely
R&S
®
FSVA/FSV
158
Quick Start Guide 1321.3066.02 ─ 06
'--------- Selection of settings to be stored ----------------------------
CALL InstrWrite(analyzer,"MMEM:SEL:HWS ON")
'Store hardware settings
CALL InstrWrite(analyzer,"MMEM:SEL:TRAC OFF")
'Do not store any traces
CALL InstrWrite(analyzer,"MMEM:SEL:LIN:ALL OFF")
'Store only the activated limit lines
'--------- Storing on the instrument -------------------------------------
CALL InstrWrite(analyzer,"MMEM:STOR:STAT 1,'C:\R_S\Instr\user\TEST1'")
'--------- Resetting the instrument --------------------------------------
CALL InstrWrite(analyzer,"*RST")
END SUB
REM ***********************************************************************
Loading Instrument Settings
In the following example, the
TEST1
data record stored under
C:\R_S\Instr\user
is loaded by the instrument:
REM ************************************************************************
Public Sub LoadSettings()
'This subroutine loads the TEST1 data record in the directory
'C:\R_S\Instr\user.
'--------- Default setting of the status register --------------------------
CALL SetupStatusReg 'Configure status register
'--------- Loading the data record -----------------------------------------
CALL InstrWrite(analyzer,"MMEM:LOAD:STAT 1,'C:\R_S\Instr\user\TEST1'")
'--------- Perform measurement using loaded data record --------------------
CALL InstrWrite(analyzer,"DISP:TRAC1:MODE WRIT")
'Set trace to Clr/Write
CALL InstrWrite(analyzer,"INIT;*WAI") 'Start sweep
END SUB
REM ************************************************************************
Setting the Data Record for Startup Recall
In the following example, the first step is to change the R&S
FSVA/FSV to the
default state. In the next step, the
TEST1
data record stored under
C:\R_S\Instr\user
is selected for the startup recall function, i.e. the data
record is then set after each
*RST
, presetting and each time the instrument is
started. For demonstration purposes, the command
*RST
is carried out again.
REM ************************************************************************
Public Sub StartupRecallSettings()
'--------- Resetting the R&S FSV ------------------------------------------
CALL InstrWrite(analyzer,"*RST")
Brief Introduction to Remote Control