Controlling the R&S
FSVA/FSV Remotely
R&S
®
FSVA/FSV
173
Quick Start Guide 1321.3066.02 ─ 06
The instrument home page displays the device information required by the LXI
standard including the VISA resource string in read-only format.
► Press the "Device Indicator" button on the "Instrument Home Page" to activate
or deactivate the LXI status icon on the status bar of the R&S
FSVA/FSV. A
green LXI status symbol indicates that a LAN connection has been estab-
lished; a red symbol indicates an error, for example, that no LAN cable is con-
nected. When a device is connecting to the instrument, the LXI logo blinks.
The "Device Indicator" setting is not password-protected.
The most important control elements in the navigation pane of the browser inter-
face are the following:
●
"LAN Configuration" opens the menu with configuration pages.
●
"Status" displays information about the LXI status of the instrument.
●
"Help > Glossary" opens a document with a glossary of terms related to the
LXI standard.
8.7.1
LAN Configuration
The LAN configuration consists of three parts:
●
"IP configuration" provides all mandatory LAN parameters.
●
"Advanced LAN Configuration" provides LAN settings that are not declared
mandatory by the LXI standard.
●
"Ping Client" provides the ping utility to verify the connection between the
instrument and other devices.
8.7.1.1
IP Configuration
The "LAN Configuration > IP configuration" web page displays all mandatory LAN
parameters and allows their modification.
The "TCP/IP Mode" configuration field controls how the IP address for the instru-
ment gets assigned (see also
Chapter 8.6.2, "Assigning the IP Address"
For the manual configuration mode, the static IP address, subnet mask, and
default gateway are used to configure the LAN. The automatic configuration
How to Configure the LAN Using the LXI Web Browser Interface