![Rohde & Schwarz 1307.9002K03 Quick Start Manual Download Page 131](http://html1.mh-extra.com/html/rohde-and-schwarz/1307-9002k03/1307-9002k03_quick-start-manual_1477765131.webp)
Basic Measurement Examples
R&S
®
FSVA/FSV
116
Quick Start Guide 1321.3066.02 ─ 06
Table 7-5: Signal generator settings (e.g. R&S SMU)
Frequency
890 MHz
Level
0 dBm
Modulation
GSM, one timeslot activated
1. Set the signal analyzer to the default state by pressing the PRESET key.
The R&S
FSVA/FSV is set to its default state.
2. Set the center frequency to 890 MHz, the span to 0 Hz and the resolution
bandwidth to 1 MHz.
a) Press the FREQ key and enter
890 MHz
.
b) Press the SPAN key and enter
0 Hz
, or press the "Zero Span" softkey.
3. Set the reference level of the R&S
FSVA/FSV to 10 dBm (= level of the signal
gen10 dB) and set the attenuation to 20 dB.
a) Press the AMPT key.
b) Enter
10 dBm
.
c) Press the "Rf Atten Manual" softkey.
d) Enter
20 dB
.
4. Set the sweep time to 1 ms.
a) Press the SWEEP key.
b) Press the "Sweeptime Manual" softkey and enter
1 ms
.
The R&S
FSVA/FSV shows the GSM burst continuously across the dis-
play.
5. By using the video trigger, set triggering on the rising edge of the burst.
a) Press the TRIG key.
b) Press the "Trg/Gate Source" softkey and select "Video" using the arrow
keys.
c) Press the "Trg/Gate Level" softkey and enter
70%
.
The R&S
FSVA/FSV shows a static image with the GSM burst at the start
of the trace. The trigger level is displayed as a horizontal line labeled with
the absolute level for the trigger threshold in the measurement diagram.
6. Configure power measurement in zero span.
a) Press the MEAS key.
b) Press the "Time Domain Power" softkey to open the submenu.
c) Switch the "Limits" softkey to "On."
d) Press the "Left Limit" softkey.
Measurements in Zero Span