![Rohde & Schwarz 1307.9002K03 Quick Start Manual Download Page 179](http://html1.mh-extra.com/html/rohde-and-schwarz/1307-9002k03/1307-9002k03_quick-start-manual_1477765179.webp)
Controlling the R&S
FSVA/FSV Remotely
R&S
®
FSVA/FSV
164
Quick Start Guide 1321.3066.02 ─ 06
●
Uses a single IANA registered port (4880), which simplifies the configuration
of firewalls
●
Supports simultaneous access of multiple users by providing versatile locking
mechanisms
●
Usable for IPv6 or IPv4 networks
Using VXI-11, each operation is blocked until a VXI-11 device handshake
returns. However, using HiSLIP, data is sent to the device using the "fire
and forget" method with immediate return. Thus, a successful return of a
VISA operation such as
viWrite()
does not guarantee that the instru-
ment has finished or started the requested command, but is delivered to the
TCP/IP buffers.
For more information see also the application note:
1MA208: Fast Remote Instrument Control with HiSLIP
8.5
Configuring the GPIB Interface
A GPIB interface is integrated on the rear panel of the instrument. You can set
the GPIB address and the ID response string. The GPIB language is set as SCPI
by default and cannot be changed for the R&S
FSVA/FSV.
For details on the GPIB interface see
Chapter 3.2.1.7, "GPIB interface"
Displaying the GPIB submenu
1. Press the SETUP key.
2. Press the "General Setup" softkey.
3. Press the "GPIB" softkey.
The submenu for setting the parameters of the remote control interface is dis-
played.
Setting the GPIB address
► In the "GPIB" menu, press the "GPIB Address" softkey.
Configuring the GPIB Interface