Meter Setup •
4
71
STAT Test Configuration
Quality control tests ensure consistent quality and accu-
rate measurements. In emergency situations, however, it
may be necessary to perform a test without delay. To
override an active lockout
1
, you can allow for
STAT Tests
(STAT =
S
hort
T
urn
A
round
T
ime).
You can set the number of tests (up to a maximum of
nine) permitted beyond the lockout. Once the number of
STAT Tests
allowed has been reached, additional tests
cannot be performed until you have successfully com-
pleted a quality control test. The default setting is
three
STAT tests.
1.
QC, operator, or new lot lockout
The results of STAT tests are always identified by the
label
STAT Test