8
• Extended Functionalities
116
■
The
Patient ID
is set to “hidden” (only possible with
a data management system), and there is a list
available:
When you start the test, the screen is blank and
Scan
is displayed.
– You may now scan a
Patient ID
using the barcode
scanner. If the patient ID is on the list, the test will
proceed. If the patient ID is not on the list, you will
get an error message.
Patient list validation
Depending on DMS settings, you have the following
options for patient list validation when working with
Patient lists:
■
A Patient ID does not have to be on the list to be
used (only valid for validation mode “off”).
■
The meter displays a notification if the Patient ID
entered is not on the list and validation mode
“notify” was selected in the DMS.
■
The meter displays an error message if the ID was
not found on the list of valid Patient IDs and valida-
tion mode “required” was selected in the DMS. To
proceed with the test, you must scan (if ID set to
“hidden”) or select (if ID set to Required) a valid
Patient ID.
■
If the “list” feature is set to be hidden, only Patient
IDs which are part of the patient list (previously
transferred from the DMS to the meter) can be
selected. It is not possible to add new Patient IDs
via scanner or keypad.
Note that the scanned
Patient ID
must already be on the
list that is sent from the DMS to the meter. You cannot
use the scanner to scan and enter a new
Patient ID
if
the
Patient ID
is set to “hidden”.
Summary of Contents for CoaguChek XS Pro
Page 1: ...CoaguChek XS Pro P O I N T O F C A R E T E S T I N G Operator s Manual ...
Page 3: ...0 6950540001 02 2015 12 EN CAN Operator s Manual Version 5 0 CoaguChek XS Pro ...
Page 8: ...8 This page intentionally left blank ...
Page 21: ...The CoaguChek XS Pro Meter 2 21 2 The CoaguChek XS Pro Meter A B C D G I J F H K L E ...
Page 110: ...7 Review Results 110 This page intentionally left blank ...
Page 134: ...9 Maintenance and Care 134 This page intentionally left blank ...
Page 138: ...10 Troubleshooting 138 This page intentionally left blank ...
Page 146: ...A Appendix 146 Codabar Code 39 Code 93 Code 128 EAN 13 Interleaved 2 5 ...
Page 150: ...B Appendix 150 This page intentionally left blank ...
Page 154: ...154 This page intentionally left blank ...
Page 155: ......