9
• Maintenance and Care
126
3
Disinfect the exterior
Take a fresh Sani-Cloth
®
Plus Germicidal Cloth from the
container. Remove excess moisture from the cloth. With
the meter powered off, wipe the meter’s exterior again.
Do not let liquid accumulate near any opening. Carefully
wipe around the cover of the test strip guide and other
openings. Make sure that no liquid enters the meter.
To properly disinfect, let the meter sit wet for a
contact time of at least 2 minutes, or as instructed
by the manufacturer in the product labeling.
Always refer to the manufacturer’s labeling for
contact time recommendations.
4
Dry the exterior
After the required contact time, dry the housing
thoroughly with a dry lint-free cloth.
Visually verify that no residual moisture is seen anywhere
on the meter at the completion of cleaning and disinfect-
ing. Ensure that the meter is completely dry before per-
forming the next patient test.
Summary of Contents for CoaguChek XS Pro
Page 1: ...CoaguChek XS Pro P O I N T O F C A R E T E S T I N G Operator s Manual ...
Page 3: ...0 6950540001 02 2015 12 EN CAN Operator s Manual Version 5 0 CoaguChek XS Pro ...
Page 8: ...8 This page intentionally left blank ...
Page 21: ...The CoaguChek XS Pro Meter 2 21 2 The CoaguChek XS Pro Meter A B C D G I J F H K L E ...
Page 110: ...7 Review Results 110 This page intentionally left blank ...
Page 134: ...9 Maintenance and Care 134 This page intentionally left blank ...
Page 138: ...10 Troubleshooting 138 This page intentionally left blank ...
Page 146: ...A Appendix 146 Codabar Code 39 Code 93 Code 128 EAN 13 Interleaved 2 5 ...
Page 150: ...B Appendix 150 This page intentionally left blank ...
Page 154: ...154 This page intentionally left blank ...
Page 155: ......