Maintenance and Care •
9
129
6
Dry the interior test strip guide
After the required contact time, carefully dry the test strip
guide with a dry lint-free cloth.
Let the inside of the test strip guide dry for at least 10
minutes before re-attaching the test strip guide cover and
testing again.
Visually verify that no residual moisture is seen anywhere
on the test strip guide and cover at the completion of
cleaning and disinfecting.
Ensure that the test strip guide and cover are completely
dry before assembling them.
7
Re-attach the test strip guide cover
Close the cover and make sure it snaps into place.