10
• Troubleshooting
136
Additional information on error
E-406
Power the meter off and remove the test strip. Repeat the
measurement using a new test strip and blood taken from
a new puncture site at the tip of another finger. Do not
touch or remove the test strip when a test is in progress.
Note on error E-406:
The CoaguChek XS PT Test strip
may be used for patients under a combination therapy of
oral anticoagulants plus heparin injections. For maximum
heparin concentrations which do not interfere with the
test, refer to the package insert. Under no circumstances,
however, should heparinized capillary tubes be used for
sample application. If capillary tubes are used, use only
the dedicated CoaguChek capillary tubes. Be sure to
apply the blood drop to the test strip within
15 seconds
of lancing the fingertip. In rare cases, patients with long
clotting times (> 8 INR; < 5% Quick) may receive an
“E-406” message on the meter display. If this error mes-
sage appears again when the test is repeated, the result
must be checked using another method.
E-406: Sample Error
Error message
Touch
to exit this message. Perform the suggested
step(s) to solve the problem.
Sample Error
Unsuitable sample
composition. Check your
procedure and repeat the
test with a fresh sample.
Check medication regimen
of patient.
E-406
Summary of Contents for CoaguChek XS Pro
Page 1: ...CoaguChek XS Pro P O I N T O F C A R E T E S T I N G Operator s Manual ...
Page 3: ...0 6950540001 02 2015 12 EN CAN Operator s Manual Version 5 0 CoaguChek XS Pro ...
Page 8: ...8 This page intentionally left blank ...
Page 21: ...The CoaguChek XS Pro Meter 2 21 2 The CoaguChek XS Pro Meter A B C D G I J F H K L E ...
Page 110: ...7 Review Results 110 This page intentionally left blank ...
Page 134: ...9 Maintenance and Care 134 This page intentionally left blank ...
Page 138: ...10 Troubleshooting 138 This page intentionally left blank ...
Page 146: ...A Appendix 146 Codabar Code 39 Code 93 Code 128 EAN 13 Interleaved 2 5 ...
Page 150: ...B Appendix 150 This page intentionally left blank ...
Page 154: ...154 This page intentionally left blank ...
Page 155: ......