5
• Testing a Blood Sample
94
STAT tests
STAT tests are a limited number of tests that can be per-
formed in emergency situations, see page 71. If the meter
is configured to perform STAT tests, you have the option
of performing a test by choosing STAT even though a QC
Lockout is in place.
1
To perform the measurement
without
performing a
quality control test, touch
Patient Test
:
2
Touch
to
confirm
the
number
of
remaining
STAT
tests.
3
Perform the test.
When a STAT test is performed, this information is stored
with the result. The number of STAT tests allowed is
reduced by 1. After all pending quality control tests are
performed, the specified number of STAT tests is availa-
ble again in case of a new lockout.
The
default setting is three
for the number of STAT
tests allowed when in QC Lockout.
Main Menu
09:15 am
04/19/2012
Control Test
Review Results
Setup
Patient Test
Logout
QC!
STAT Info
3
STAT tests remaining
before lockout.
I-011
Test
09:15 am
04/19/2012
P-ID: PID01
Code: 184
2.5
INR
Smith, Henry
Op:
Operator 1
Par:
PT
STAT Test
21
%Q
04/19/2012 09:14 am
Summary of Contents for CoaguChek XS Pro
Page 1: ...CoaguChek XS Pro P O I N T O F C A R E T E S T I N G Operator s Manual ...
Page 3: ...0 6950540001 02 2015 12 EN CAN Operator s Manual Version 5 0 CoaguChek XS Pro ...
Page 8: ...8 This page intentionally left blank ...
Page 21: ...The CoaguChek XS Pro Meter 2 21 2 The CoaguChek XS Pro Meter A B C D G I J F H K L E ...
Page 110: ...7 Review Results 110 This page intentionally left blank ...
Page 134: ...9 Maintenance and Care 134 This page intentionally left blank ...
Page 138: ...10 Troubleshooting 138 This page intentionally left blank ...
Page 146: ...A Appendix 146 Codabar Code 39 Code 93 Code 128 EAN 13 Interleaved 2 5 ...
Page 150: ...B Appendix 150 This page intentionally left blank ...
Page 154: ...154 This page intentionally left blank ...
Page 155: ......