120
5.3
Performing a liquid quality control test
1
Place the meter on a level, vibration-free surface or
hold it in your hand so it is roughly horizontal.
2
Power the meter on by pressing
.
You can also power on the meter directly by inserting a
test strip or connecting the power adapter.
3
Wait until the Main Menu is displayed, or log on as
4
Check the battery level.
5
Check that the date and time are correct. Correct
any wrong entries as described in Chapter 3,
■
If the battery icon turns red (one bar left), there may
not be enough power left for another test.
■
If there are no bars left in the battery icon, you cannot
perform any more tests. Power the meter off by
pressing .
In both cases, restore power by recharging the battery
pack, or using the power adapter.
Main Menu
09:15 am
05/29/2015
Control Test
Review Results
Setup
Patient Test
Summary of Contents for CoaguCheck Pro II
Page 1: ...CoaguChek Pro II Operator s Manual ...
Page 3: ...0 7154879001 02 2015 11 EN CoaguChek Pro II Operator s Manual Version 2 0 ...
Page 6: ...6 This page intentionally left blank ...
Page 28: ...28 1 3 Overview of the meter elements C D F H I G J B A E E ...
Page 42: ...42 This page intentionally left blank ...
Page 116: ...116 This page intentionally left blank ...
Page 138: ...138 This page intentionally left blank ...
Page 168: ...168 Codabar Code 39 Code 93 Code 128 EAN 13 Interleaved 2 5 PDF417 QR Code DataMatrix ...
Page 170: ...170 This page intentionally left blank ...
Page 174: ...174 This page intentionally left blank ...
Page 182: ...182 This page intentionally left blank ...
Page 186: ...186 This page intentionally left blank ...
Page 187: ......