118
5.2
Preparing to perform a liquid quality control test
Prepare for a liquid quality control test in the same way
you would prepare to perform a test with a capillary blood
sample. The only difference is the use of control solution
instead of blood.
1
Have the test strip container at hand.
2
If you are using the test strip lot for the first time, make
sure that the code chip that came with these test
strips is at hand.
3
Make sure the bottle of freeze-dried (lyophilized) con-
trol plasma and the dropper for making the control
solution are at hand. This bottle should remain refrig-
erated (not frozen) until use.
4
Make sure that the quality control code chip that
came with the control solution is at hand.
5
Open the lid of the bottle and remove the rubber cap.
6
Hold the dropper with the sealed dropper neck point-
ing upward, then cut off the end of the cap with scis-
sors. Do not hold the dropper close to your face.
To avoid loss of diluent, hold the dropper by the stem;
do not squeeze the bulb of the dropper while cutting
the tip.
Summary of Contents for CoaguCheck Pro II
Page 1: ...CoaguChek Pro II Operator s Manual ...
Page 3: ...0 7154879001 02 2015 11 EN CoaguChek Pro II Operator s Manual Version 2 0 ...
Page 6: ...6 This page intentionally left blank ...
Page 28: ...28 1 3 Overview of the meter elements C D F H I G J B A E E ...
Page 42: ...42 This page intentionally left blank ...
Page 116: ...116 This page intentionally left blank ...
Page 138: ...138 This page intentionally left blank ...
Page 168: ...168 Codabar Code 39 Code 93 Code 128 EAN 13 Interleaved 2 5 PDF417 QR Code DataMatrix ...
Page 170: ...170 This page intentionally left blank ...
Page 174: ...174 This page intentionally left blank ...
Page 182: ...182 This page intentionally left blank ...
Page 186: ...186 This page intentionally left blank ...
Page 187: ......