![Roche CoaguCheck Pro II Operator'S Manual Download Page 88](http://html1.mh-extra.com/html/roche/coagucheck-pro-ii/coagucheck-pro-ii_operators-manual_1472337088.webp)
88
STAT Test Configuration
Quality control tests ensure consistent quality and accu-
rate measurements. In emergency situations, however, it
may be necessary to perform a test without delay. To
override an active lockout
1
, you can allow
STAT Tests
(STAT =
S
hort
T
urn
A
round
T
ime).
You can set the number of tests (up to nine) permitted
beyond the lockout. Once the number of
STAT Tests
allowed has been reached, additional tests cannot be
performed until you have successfully completed a qual-
ity control test. The default setting is
three
STAT tests.
1.
QC, operator, or new lot lockout
The results of STAT tests are always identified by the
label
STAT Test
when they are displayed (see page 115).
The number of STAT tests performed is counted per test
parameter used. Once the maximum number of STAT
tests has been reached, this parameter cannot be
tested without performing a control test first. Other test
parameters that have not yet reached this limit are still
available.
Summary of Contents for CoaguCheck Pro II
Page 1: ...CoaguChek Pro II Operator s Manual ...
Page 3: ...0 7154879001 02 2015 11 EN CoaguChek Pro II Operator s Manual Version 2 0 ...
Page 6: ...6 This page intentionally left blank ...
Page 28: ...28 1 3 Overview of the meter elements C D F H I G J B A E E ...
Page 42: ...42 This page intentionally left blank ...
Page 116: ...116 This page intentionally left blank ...
Page 138: ...138 This page intentionally left blank ...
Page 168: ...168 Codabar Code 39 Code 93 Code 128 EAN 13 Interleaved 2 5 PDF417 QR Code DataMatrix ...
Page 170: ...170 This page intentionally left blank ...
Page 174: ...174 This page intentionally left blank ...
Page 182: ...182 This page intentionally left blank ...
Page 186: ...186 This page intentionally left blank ...
Page 187: ......