Testing a Blood Sample •
4
107
If you use a new test strip lot and have not inserted the
code chip yet, you must do so now. Otherwise you cannot
perform a test. The meter displays the number of the
code chip belonging to the new test strip lot. Depending
on the setting, you may also be required to run a liquid
quality control test at this point.
The hourglass icon shows that the test strip is warming
up. When the warming-up process is complete, a further
beep (provided the beeper is enabled) indicates that you
can now apply blood.
The blood drop icon flashes to indicate that the meter is
ready to perform the test and is waiting for blood to be
applied.
A countdown begins. You must apply the drop of blood to
the test strip before the countdown ends. Otherwise you
will receive an error message.
When the test is running, no operator action is required.
The display dims to save energy.
Test
09:15 am
05/29/2015
184
P-ID:
PID01
Smith, Henry
Op:
Operator 1
Par:
PT
Test
09:15 am
05/29/2015
P-ID:
PID01
Code:
184
Smith, Henry
Op:
Operator 1
Par:
PT
Test
09:15 am
05/29/2015
P-ID:
PID01
Code:
184
3:00 Min.
Smith, Henry
Op:
Operator 1
Par:
PT
Summary of Contents for CoaguCheck Pro II
Page 1: ...CoaguChek Pro II Operator s Manual ...
Page 3: ...0 7154879001 02 2015 11 EN CoaguChek Pro II Operator s Manual Version 2 0 ...
Page 6: ...6 This page intentionally left blank ...
Page 28: ...28 1 3 Overview of the meter elements C D F H I G J B A E E ...
Page 42: ...42 This page intentionally left blank ...
Page 116: ...116 This page intentionally left blank ...
Page 138: ...138 This page intentionally left blank ...
Page 168: ...168 Codabar Code 39 Code 93 Code 128 EAN 13 Interleaved 2 5 PDF417 QR Code DataMatrix ...
Page 170: ...170 This page intentionally left blank ...
Page 174: ...174 This page intentionally left blank ...
Page 182: ...182 This page intentionally left blank ...
Page 186: ...186 This page intentionally left blank ...
Page 187: ......