Testing a Blood Sample •
4
111
When interpreting results, refer to the detailed informa-
tion on limitations and interferences included in the limi-
tations section of the test strip package insert.
If a “C” is displayed along with the result:
This may occur if the hematocrit value is very low or due
to erroneous blood collection (e.g., wet hands). Repeat
the test. If using capillary blood, make sure that the
patient's hands are dry. If the message persists, perform
a hematocrit check.
If a “
*
” is displayed along with the result:
This may occur if the test result is out of normal range.
Touch the
*
icon to display the
Out Of Normal Range
information screen.
Test
09:15 am
05/29/2015
P-ID:
PID05
Code:
208
1.6
INR
Müller, David
Op:
Schulz
Par:
PT
05/29/2015 09:14 am
C
Test
09:15 am
05/29/2015
P-ID:
PID05
Code:
208
1.6
INR
Müller, David
Op:
Schulz
Par:
PT
05/29/2015 09:14 am
C
Summary of Contents for CoaguCheck Pro II
Page 1: ...CoaguChek Pro II Operator s Manual ...
Page 3: ...0 7154879001 02 2015 11 EN CoaguChek Pro II Operator s Manual Version 2 0 ...
Page 6: ...6 This page intentionally left blank ...
Page 28: ...28 1 3 Overview of the meter elements C D F H I G J B A E E ...
Page 42: ...42 This page intentionally left blank ...
Page 116: ...116 This page intentionally left blank ...
Page 138: ...138 This page intentionally left blank ...
Page 168: ...168 Codabar Code 39 Code 93 Code 128 EAN 13 Interleaved 2 5 PDF417 QR Code DataMatrix ...
Page 170: ...170 This page intentionally left blank ...
Page 174: ...174 This page intentionally left blank ...
Page 182: ...182 This page intentionally left blank ...
Page 186: ...186 This page intentionally left blank ...
Page 187: ......