![Roche CoaguCheck Pro II Operator'S Manual Download Page 132](http://html1.mh-extra.com/html/roche/coagucheck-pro-ii/coagucheck-pro-ii_operators-manual_1472337132.webp)
132
12
Apply a single drop of the sample directly from the
dropper to the semicircular, transparent sample
application area on top of the test strip. Do not add
more of the sample.
You hear a beep when you have applied enough of the
sample (provided the beeper is enabled). The dropper
icon disappears and the test starts.
The result of the proficiency test is displayed, the com-
ment
Proficiency
is automatically added. If you want to
add further comments, touch
. The test result is auto-
matically saved to memory.
After the test result is displayed, touch
. You will be
prompted to remove the strip.
13
Remove the test strip from the meter.
14
Power the meter off.
15
Clean the meter if this becomes necessary (see
).
Profi ciency
09:15 am
05/29/2015
QC:
202
Code:
101
Par:
aPTT
Profi ciency
09:15 am
05/29/2015
QC:
202
Code:
101
SEC
06/03/2015 11:30 am
Par:
aPTT
Profi ciency
51.6
Dispose of used test strips from proficiency testing in
accordance with the disposal policy of your facility.
Summary of Contents for CoaguCheck Pro II
Page 1: ...CoaguChek Pro II Operator s Manual ...
Page 3: ...0 7154879001 02 2015 11 EN CoaguChek Pro II Operator s Manual Version 2 0 ...
Page 6: ...6 This page intentionally left blank ...
Page 28: ...28 1 3 Overview of the meter elements C D F H I G J B A E E ...
Page 42: ...42 This page intentionally left blank ...
Page 116: ...116 This page intentionally left blank ...
Page 138: ...138 This page intentionally left blank ...
Page 168: ...168 Codabar Code 39 Code 93 Code 128 EAN 13 Interleaved 2 5 PDF417 QR Code DataMatrix ...
Page 170: ...170 This page intentionally left blank ...
Page 174: ...174 This page intentionally left blank ...
Page 182: ...182 This page intentionally left blank ...
Page 186: ...186 This page intentionally left blank ...
Page 187: ......