6
• Storing Test Strip, Control Solution, and Linearity Solution Information in the Meter
78
The parameters for control solutions consist of four sepa-
rate values.
3
Use the keypad to enter the desired values one
after another:
– Minimum limit value for Level 1
– Maximum limit value for Level 1
– Minimum limit value for Level 2
– Maximum limit value for Level 2
4
Touch
to confirm each separate entry and
continue to the next entry.
1
2
3
4
5
6
7
8
9
0
Control L2 Max
19.6
mmol/L
12:48 pm
09/18/13
.
1
2
3
4
5
6
7
8
9
0
Control L2 Min
14.5
mmol/L
12:48 pm
09/18/13
.
1
2
3
4
5
6
7
8
9
0
Strip Expiration
11/30/13
12:48 pm
09/18/13