10
• Maintenance and Care
136
Cleaning the barcode scanner
window
The barcode scanner window should be cleaned periodi-
cally. Use a clean, dry cloth to wipe the barcode scanner
window.
Cleaning/disinfecting the base
unit
1
Unplug the base unit before cleaning/disinfecting.
2
Wipe over the surfaces with a soft lint-free cloth
slightly dampened (wring out any excess).
If using commercially available pre-moistened cleaning/
disinfecting wipes,
squeeze off excess solution
or blot
on a dry paper towel to remove any excess solution
before wiping the surface of the base unit.
Do not
wipe the electrical connectors on the back of the
base unit.
3
Dry
the base unit thoroughly after cleaning/disin-
fecting. Visually verify that no solution is seen in the
base unit connectors at the completion of disin-
fecting. If fluid is allowed to collect in any connector,
severe damage can occur to the meter and the base
unit.
4
Plug in the base unit.
NOTICE
Ensure meter/base unit (including connectors) are
thoroughly dried after cleaning or disinfecting. A flashing
LED (red) on the base unit indicates a fault condition.
NOTICE
Do not spray the base unit directly with solutions as this
could cause the solution to enter the case and damage
the electronic components.
If cleaning/disinfecting solution does get on the connec-
tors or pools in the base unit, unplug the base unit, then
dry the components
with a dry cloth or gauze pad
before returning the meter to the base unit.
Summary of Contents for ACCU-CHEK Inform II
Page 1: ...Accu Chek Inform II BLOOD GLUCOSE MONITORING SYSTEM Operator s Manual ...
Page 3: ...0 4807839001 05 2013 09 EN Operator s Manual Version 5 0 A ACCU CCU CHEK HEK Inform II System ...
Page 6: ...6 This page intentionally left blank ...
Page 12: ...12 This page intentionally left blank ...
Page 104: ...8 Proficiency Testing 104 This page intentionally left blank ...
Page 160: ...12 General Product Information 160 This page intentionally left blank ...
Page 206: ...D Supplement for Observed Test Sequence 206 This page intentionally left blank ...
Page 210: ...Index 210 This page intentionally left blank ...
Page 211: ...211 ...