Appendix •
A
167
QC algorithm
QC Algorithm
0: None (always OK)
1: Last result OK
2: Time of day (DMS only)
3: Shift (DMS only)
4: Hours
5: Strip count
6: Time of day rotate
(DMS only)
7: Shift rotate (DMS only)
8: Hours rotate (DMS
only)
9: Strip count rotate
(DMS only)
0
(Y)
Y
QC hours (Number of hours between
required control measurements if QC algo-
rithm is set to “Hours” or “hours rotate”)
0 – 9,999 hours
24
Y
Y
QC shift length (a PASS control remains
valid for twice this amount in time after
shift start if the QC algorithm is set to
“Shift” or “Shift rotate”)
1, 2, 3, 4, 6, 8, 12, 24 hours 8
N
Y
QC shift start (time of the shift start if the
QC algorithm is set to “Shift” or “Shift
rotate”)
0 – 23 hours
6
N
Y
QC strip count (number of tests a PASS
control remains valid)
1 – 999
50
Y
Y
QC time of day
6 POCT1-A data fields
00:00 – 23:59
06:00
09:00
12:00
15:00
18:00
22:00
N
Y
QC time of day set
6 POCT1-A data fields
0 or 1
1 0 0 0 0 0
N
Y
Subject/Attribute
Range
Default
Device
DMS
Summary of Contents for ACCU-CHEK Inform II
Page 1: ...Accu Chek Inform II BLOOD GLUCOSE MONITORING SYSTEM Operator s Manual ...
Page 3: ...0 4807839001 05 2013 09 EN Operator s Manual Version 5 0 A ACCU CCU CHEK HEK Inform II System ...
Page 6: ...6 This page intentionally left blank ...
Page 12: ...12 This page intentionally left blank ...
Page 104: ...8 Proficiency Testing 104 This page intentionally left blank ...
Page 160: ...12 General Product Information 160 This page intentionally left blank ...
Page 206: ...D Supplement for Observed Test Sequence 206 This page intentionally left blank ...
Page 210: ...Index 210 This page intentionally left blank ...
Page 211: ...211 ...