Maintenance and Care •
10
137
Cleaning/disinfecting the
accessory box
For cleaning,
you may wipe the surfaces with a soft
cloth slightly dampened (not wet) with 70 % (or less)
solution of isopropyl alcohol in water or with 70 %
isopropyl alcohol, full strength.
For disinfection:
Acceptable active ingredients for
disinfecting the accessory box are:
■
0.625 % (or less) solution of sodium hypochlorite in
water (bleach).
1
Wipe the surface with a soft cloth slightly damp-
ened (not wet).
If spraying the accessory box, place it on a flat sur-
face or table. Ensure that it is completely empty.
2
Allow the accessory box to air dry for the recom-
mended contact time according to the cleaning/
disinfecting solution product labeling. Do not place
the meter in the accessory box until all cleaning/
disinfecting steps are completed.
3
Dry the accessory box thoroughly with a dry cloth
or gauze and visually verify that no solution is seen
in the accessory box at the completion of cleaning/
disinfecting.
Ensure the accessory box is thoroughly dried before
using or filling it again.
NOTICE
Use of disinfectants containing other active ingredients
could result in damage to the accessory box.
Summary of Contents for ACCU-CHEK Inform II
Page 1: ...Accu Chek Inform II BLOOD GLUCOSE MONITORING SYSTEM Operator s Manual ...
Page 3: ...0 4807839001 05 2013 09 EN Operator s Manual Version 5 0 A ACCU CCU CHEK HEK Inform II System ...
Page 6: ...6 This page intentionally left blank ...
Page 12: ...12 This page intentionally left blank ...
Page 104: ...8 Proficiency Testing 104 This page intentionally left blank ...
Page 160: ...12 General Product Information 160 This page intentionally left blank ...
Page 206: ...D Supplement for Observed Test Sequence 206 This page intentionally left blank ...
Page 210: ...Index 210 This page intentionally left blank ...
Page 211: ...211 ...