1 ABOUT THIS MANUAL
Manual ERZ2000-DI · EN1.2 · 2017, December 21st
17
1.5.2
Seal diagram for devices with MID approval
The ERZ2000-DI leaves the company with a setting agreed upon with the cus-
tomer, which is protected from "significant" modifications by seals and metrological
sealing, or it leaves the factory without seals. In the latter case, a calibration official
must provide the device with seals according to the desired settings in accordance
with the seal diagram. The metrological seals must also be applied, as specified in
the seal diagram.
Figure 2: Seal diagram - front
and
Figure 4: Seal diagram - rear
show the specified positions of the seals.
Caution
The ERZ2000-DI must only be used for custody transfer with unbroken
seal. Removal or damage to seals normally entails considerable expenses!
Re-application of seals must only be carried out by an officially recog-
nized inspection authority or calibration officials!
Figure 2: Seal diagram - front
FLOW COMPUTER SYSTEM ERZ 2000-DI
Type plate for:
ERZ2004-DI/
ERZ2104-DI
Optional type plate
for maximum load
Metrological seal
Seal
Frontal view