42
Operation Modes
Measurement lower limit
The lower limit for measurement is determined by the noise level.
This is the level that is indicated when a dummy load of 1000 pF is connected
to the PICKUP connector and the sensitivity switch is set to 5.00 pC/(m/s
2
).
In order to limit the infl uence of noise upon the measurement to no more
than 1 dB (approx. 10%), the lower limit is set to a point two times higher
than the noise level.
Input
Acceleration/Velocity/Displacement
HPF/LPF
Inherent noise
(20°C 40% RH)
1000 pF dummy
Acceleration
OFF/OFF
0.01 m/s
2
(RMS)
Velocity
OFF/OFF
0.1 mm/s
(RMS)
Displacement
OFF/OFF
0.0015 mm
(RMS)
PV-87
Acceleration
OFF/OFF
0.0013 m/s
2
(RMS)
Velocity
OFF/OFF
0.013 mm/s
(RMS)
Displacement
OFF/OFF
0.00025 mm (RMS)
PV-85
Acceleration
OFF/OFF
0.008 m/s
2
(RMS)
Velocity
OFF/OFF
0.08 mm/s
(RMS)
Displacement
OFF/OFF
0.0013 mm
(RMS)
PV-90B
Acceleration
OFF/OFF
0.25 m/s
2
(RMS)
Velocity
OFF/OFF
2.5 mm/s
(RMS)
Displacement
OFF/OFF
0.05 mm
(RMS)
PV-91C
*1
Acceleration
OFF/OFF
0.02 m/s
2
(RMS)
Velocity
OFF/OFF
0.53 mm/s
(RMS)
Displacement
OFF/OFF
0.024 mm
(RMS)
PV-90I
*2
Acceleration
OFF/OFF
0.007 m/s
2
(RMS)
Velocity
OFF/OFF
0.117 mm/s
(RMS)
Displacement
OFF/OFF
0.0068 mm
(RMS)
Measurement conditions: temperature 20°C, unit powered from batteries using BP-17
*1 PV-91C Inherent noise standard value :
0.02
m/s
2
(TYP.),0.04 m/s
2
(MAX.), 1 Hz to 20 kHz.
*2 PV-91CH Inherent noise standard value :
0.007 m/s
2
(TYP.), 0.009 m/s
2
(MAX.), 1 Hz to 15 kHz.
Summary of Contents for UV-15
Page 2: ......
Page 10: ...viii...