25
Accelerometer Selection and Connection
The following types of accelerometers can be connected to the UV-15.
Accelerometer type
Sensitivity setting range
Piezoelectric accelerometer
0.100 to 99.9 pC/(m/s
2
)
Accelerometer with integrated preamplifi er
0.100 to 99.9 mV/(m/s
2
)
TEDS compliant accelerometer with integrated preamplifi er
0.100 to 99.9 mV/(m/s
2
)
Important
Before connecting or disconnecting an accel-
erometer, be sure to turn off the power supply
to the system.
Major piezoelectric accelerometers from Rion (representative values)
Compact,
lightweight
General
High-
output
High-
temperature
3-axis
Model
PV-90B
PV-85
PV-87
PV-65
PV-97C
PV-93
Mass (g)
1.2
23
115
26
4.7
30
Typ. sensitivity
(pC/(m/s
2
))
0.18
6.42
40
7.14
0.12
0.831
Frequency
range (Hz)
*1
1 to 25000 1 to 7000 1 to 3000 1 to 9000
1 to 10000 (X/Y)
1 to 15000 (Z)
1 to 8000 (axis 2)
1 to 4000 (axis 1/axis 3)
Mounting resonance
frequency (kHz)
*2
70
24
9
25
---
---
Connection cable
*3
VP-51L
VP-51A
VP-51A
VP-51B
VP-51L × 3
VP-51C × 3
Major accelerometers with integrated preamplifi er from Rion (rep-
resentative values)
Compact, lightweight
Model
PV-91C
PV-91CH
Mass (g)
1.8
3
Typ. sensitivity (mV/(m/s
2
))
*1
1
11
Frequency range (Hz)
*2
1 to 20000
*4
1 to 15000
*5
Mounting resonance frequency (kHz)
*2
55
50
Connection cable
*3
VP-51LC
VP-51LC
*1 For actual sensitivity values, see sensitivity chart supplied with accelerometer.
*2 Representative value when mounted on a fl at surface using standard mounting method
*3 Supplied with accelerometer
*4 At 150°C to 170°C, 1 Hz to 2 Hz (±15%)
*5 0.6 Hz to 20 kHz, 0.5 Hz to 20 kHz (±30%)