36
Operation Modes
Measurement setup mode
This mode serves for selecting the measurement mode (unit), display char-
acteristics, HPF (high-pass fi lter), and LPF (low-pass fi lter) settings.
Be sure to make or check these settings before starting a measurement.
Measurement setup mode functions
Settings
Select measurement mode (unit)
m/s
2
(Acceleration: ACC), mm/s (Velocity: VEL), mm
(Displacement: DISP)
Display characteristics
RMS, EQ PEAK, EQ P-P
HPF setting
OFF (---), 3 Hz, 5 Hz, 10 Hz, 15 Hz, 20 Hz, 30 Hz, 50 Hz,
100 Hz, 150 Hz, 200 Hz
LPF setting
300 Hz, 500 Hz, 1 kHz, 1.5 kHz, 2 kHz, 3 kHz, 5 kHz,
10 kHz, 15 kHz, 20 kHz, OFF (---)
Setup procedure
1. Pressing the MEAS key in measurement mode activates the mea-
surement setup mode, and allows you to set the measurement mode
(unit).
You can change the setting with the RANGE keys.
m/s
2
: Acceleration
(ACC)
mm/s: Velocity
(VEL)
mm: Displacement
(DISP)
When the setting is as required, press the MEAS key.
2. When you press the MEAS key, the display characteristics setup
screen appears.
You can change the setting with the RANGE keys.
RMS: Effective
value
EQ PEAK:
Equivalent peak value (RMS ×
2
)
EQ P-P:
Equivalent peak-to-peak value (EQPEAK × 2)
When the setting is as required, press the MEAS key.
3. When you press the MEAS key, the HPF (high-pass fi lter) setup
screen appears.
You can change the setting with the RANGE keys.
OFF, 3 Hz, 5 Hz, 10 Hz, 15 Hz, 20 Hz, 30 Hz, 50 Hz, 100 Hz,
150 Hz, 200 Hz
The OFF setting is shown as "---".
When the setting is as required, press the MEAS key.