31
Operation Modes
The UV-15 has the following operation modes.
Operation mode
Function outline
Input setup mode
Make settings for input (connector, sensitivity)
and ID number
TEDS communication
mode
Receive sensitivity data and other information from
TEDS compliant accelerometer with integrated
preamplifi er
Measurement setup
mode
Make settings for acceleration, velocity, displace-
ment, display characteristics, HPF, and LPF
Calibration mode
Supply CAL signal for calibration of external
equipment
UV-22 communication
mode
Control UV-15 from connected Interface Unit
UV-22
Measurement mode
Make measurements
Check mode
Check the software version of the UV-15
Input setup mode
This serves to make settings for the connected accelerometer and select an
ID number for the UV-15. Be sure to make or check these settings before
starting a measurement.
Input setup mode functions
Item
Settings
Input selection
FRONT CHARGE
FRONT CCLD
FRONT CCLD TEDS
REAR
Sensitivity
0.100 to 0.999 in 0.001 steps
1.00 to 9.99 in 0.01 steps
10.0 to 99.9 in 0.1 steps
ID number
1 to 16