34
Operation Modes
TEDS communication mode
In this mode, the UV-15 communicates with the TEDS sensor to receive
sensitivity information and set the sensitivity accordingly.
TEDS communication can also be carried out in the input setup mode. If a
TEDS sensor was used previously, TEDS communication will also be carried
out when power is supplied the next time.
TEDS sensor support
The following types of TEDS sensors are supported.
Template ID: 24, 25
IEEE 1451.4-2000
REMOTE
FRONT
REAR
ACOUST
F
S
10ms
CHARGE
CCLD TEDS
No.
SENS
dB
REMOTE
FRONT
REAR
ACOUST
F
S
10ms
CHARGE
CCLD TEDS
No.
SENS
dB
REMOTE
FRONT
REAR
ACOUST
F
S
10ms
CHARGE
CCLD TEDS
No.
SENS
dB
Flashing
" " moves flashing
from left to right
Normal
communication
Constantly on
During TEDS communication
After TEDS communication
Error has
occurred
Flashing
After TEDS communication is completed,
sensitivity is set automatically and ID
number setup screen (page 35) appears.
Err indication
flashing
Meaning of error number
Err-1: No TEDS sensor connected
Err 0: Other error
Err 1: Template ID=27 or 25 (out of range)
Err 2: Template ID=28 or 26 (out of range)
Err 3: Template ID=12
(out of range)
Err 4: Unknown template ID
After error indication, press MEAS key once to
switch to input setup mode screen (CCLD setting)
TEDS communication
complete (error)
Shows sensitivity
as set by TEDS
communication