20
Linking
Using a single UV-15 unit
Attach the supplied link plate to the bottom of the unit as shown below, to
stabilize the unit when it is used in a free-standing confi guration.
Use the screws removed from the unit.
Why not to use other screws:
Screws too short will not properly secure the plate.
Screws too long may damage parts inside the unit.
1. Remove two screws from the cover plates on the bottom of the unit
(the two screws that are nearer to the front side). Refer to the left-side
illustration (Fig. 1) below.
2. Use the screws removed in step 1 to fasten the supplied link plate
sideways. Refer to the right-side illustration (Fig. 2) below.
PICKUP
VIBRA
TION METER UNI
T
UV
-15
RMS
EQ PEAK
EQ P-P
Fig. 1
Remove these
two screws
Unit bottom
PICKUP
VIBRA
TION METER UNI
T
UV
-15
RMS
EQ PEAK
EQ P-P
Fig. 2
Fasten link plate to unit
(pay attention to correct
orientation of flat-head
screw holes).
Supplied link plate
PICKUP
VIBRA
TION METER UNI
T
RMS
EQ PEAK
EQ P-P
UV
-1
5
Do not plug the link cable into the
connector on the unit itself. Otherwise
the system may become more
susceptible to noise and other external
interference.