Operation Modes
2. When you press the INPUT key, the unit switches to the sensitivity
(SENS) setting screen. Use the RANGE keys to make a setting in
the range from 0.100 to 0.999, 1.00 to 9.99, or 10.0 to 99.9.
With the DISPLAY key, you can change the decimal point of the
sensitivity setting. Holding down a RANGE key changes the numeric
value more quickly.
When the setting is as required, press the INPUT key.
3. When you press the INPUT key, the unit switches to the ID number
setting screen.
The ID number setting is required for unit identifi cation when multiple
UN-14 and UV-15 units are used. For details, see page 35.
The setting range is 1 to 16. Use the RANGE keys to select the
number. Holding down a RANGE key changes the numeric value
more quickly. For multi-channel confi gurations, each unit must have
a unique ID number.
When the setting is as required, press the INPUT key.
All changed settings will become active. The input setup mode is
terminated and the unit switches to measurement mode.
If you press the MEAS key at any point during steps 1 to 3, the settings will
become active and the unit switches to measurement mode immediately.
Note
About the ID number
When several UN-14 and UV-15 units are used
in a multi-channel confi guration, the ID number
is required for identifi cation.
When using the UV-22, make sure that each unit
has a unique ID number. If there is more than
one unit with the same number, communication
with the UV-22 will not function correctly.
Sensitivity setting units
pC/(m/s
2
): FRONT
CHARGE
mV/(m/s
2
): FRONT
CCLD
FRONT CCLD TEDS
REAR