
35
Operation Modes
Setting the ID number
When the UV-15 is used in conjunction with the Interface Unit UV-22, an
ID number for the UV-15 must be set fi rst.
To set the ID number, the keys on the operational panel of the respective
UV-15 unit must be used.
The default setting for the ID number is No. 1. Be sure to change the setting
when there is more than one unit. The setting range is No. 1 to No. 16.
The ID number is used to identify the unit for communication. If there is
more than one unit in the same system with the same ID number, correct
communication will not be possible.
See illustration below
1. When you press the INPUT key on the operation panel of the UV-15
three times in succession, the ID number starts to fl ash.
2. Use the RANGE keys to set the number. When the setting is complete,
press the MEAS key to return to the measurement screen.
Make the setting for each unit individually, so that each unit has a unique
ID number.
Note
When a TEDS sensor is connected and the FRONT
CCLD TEDS setting is selected, pressing the INPUT
key will automatically set the sensitivity. The ID
number setup screen then appears.
FRONT
REAR
CO S
CHARGE
CCLD TEDS
No.
ID number flashes
ID number setup screen
VIBRATION METER UNIT
UV-15
CAL
DISPLAY
INPUT
MEAS
RANGE
RMS
EQ PEAK
EQ P-P
SETUP
INPUT key
MEAS key
RANGE keys
UV-15 operation panel
Summary of Contents for UV-15
Page 2: ......
Page 10: ...viii...