![Rigol DS4000E Series Programming Manual Download Page 83](http://html1.mh-extra.com/html/rigol/ds4000e-series/ds4000e-series_programming-manual_1460996083.webp)
RIGOL
DS4000E Programming Guide
2-71
:CURSor:MANual:CAY
Syntax
:CURSor:MANual:CAY <ay>
:CURSor:MANual:CAY?
Description
Sets or queries the vertical position of Cursor A (Y cursor) in the manual mode of cursor measurement.
Parameter
Name
Type
Range
Default
<ay>
Integer
0 to 399
100
Remarks
The vertical position of the cursor is defined by the pixel coordinate of the screen. The range of the pixel
coordinate on the screen vertically is from 0 to 399 (from top to bottom).
Return Format
The query returns an integer between 0 and 399.
:CURSor:MANual:CBX
Syntax
:CURSor:MANual:CBX <bx>
:CURSor:MANual:CBX?
Description
Sets or queries the horizontal position of Cursor B (X cursor) in the manual mode of cursor measurement.
Parameter
Name
Type
Range
Default
<bx>
Integer
0 to 699
550
Remarks
The horizontal position of the cursor is defined by the pixel coordinate of the screen. The range of the pixel
coordinate on the screen horizontally is from 0 to 699 (from left to right).
Return Format
The query returns an integer between 0 and 699.
:CURSor:MANual:CBY
Syntax
:CURSor:MANual:CBY <by>
:CURSor:MANual:CBY?
Description
Sets or queries the vertical position of Cursor B (Y cursor) in the manual mode of cursor measurement.
Parameter
Name
Type
Range
Default
<by>
Integer
0 to 399
300