RIGOL
2-42
DS4000E Programming Guide
Parameter
Name
Type
Range
Default
<bool>
Bool
{{1|ON}|{0|OFF}}
0|OFF
Remarks
1|ON: turns the waveform invert on. The waveform voltage values are inverted and then displayed.
0|OFF: turns the waveform invert off. The waveform is displayed normally.
Return Format
The query returns 1 or 0.
:CALCulate:ADVanced:VARiable1
:CALCulate:ADVanced:VARiable2
Syntax
:CALCulate:ADVanced:VARiable1 <variable>
:CALCulate:ADVanced:VARiable1?
:CALCulate:ADVanced:VARiable2 <variable>
:CALCulate:ADVanced:VARiable2?
Description
Sets or queries Variable1 or Variable2 in the expression of advanced operation.
Parameter
Name
Type
Range
Default
<variable>
Real
Refer to Remarks
0
Remarks
<variable> is in the AeB format. Wherein, A is mantissa, and its available range is from -9.9999 to +9.9999
(with the minimum step of 0.0001, that is, precision to four decimals). B refers to the exponent with 10 as
the base, and its available value is an integer ranging from -9 to +9. In addition, <variable> can also be
expressed in decimal, and its available range is the same with that of the AeB format.
Return Format
The query returns Variable1 or Variable2 in scientific notation.
:CALCulate:ADVanced:VOFFset
Syntax
:CALCulate:ADVanced:VOFFset <offs>
:CALCulate:ADVanced:VOFFset?
Description
Sets or queries the vertical position of the advanced operation waveform. The unit is related to the
expression of advanced operation.
Parameter
Name
Type
Range
Default
<offs>
Real
(-40 x vertical scale of the advanced operation waveform) to
(40 x vertical scale of the advanced operation waveform)
0