RIGOL
DS4000E Programming Guide
2-195
:TRIGger:HOLDoff
Syntax
:TRIGger:HOLDoff <value>
:TRIGger:HOLDoff?
Description
Sets or queries the trigger holdoff time. The default unit is s.
Parameter
Name
Type
Range
Default
<value>
Real
100 ns to 10 s
100 ns
Remarks
The trigger types that support the holdoff include Edge trigger, Pulse trigger, Runt trigger, Slope trigger, and
Pattern trigger.
Return Format
The query returns the trigger holdoff time in scientific notation.
:TRIGger:MODE
Syntax
:TRIGger:MODE <mode>
:TRIGger:MODE?
Description
Sets or queries the trigger type.
Parameter
Name
Type
Range
Default
<mode>
Discrete
{EDGE|PULSe|RUNT|NEDG|SLOPe|VIDeo|PATTern|
RS232|IIC|SPI|CAN|FLEXray|USB}
EDGE
Return Format
The query returns EDGE, PULS, RUNT, NEDG, SLOP, VID, PATT, RS232, IIC, SPI, CAN, FLEX, or USB.
:TRIGger:STATus?
Syntax
:TRIGger:STATus?
Description
Queries the current operating state.
Remarks
The operating states include RUN, STOP, T'D (triggered), WAIT, and AUTO.
Return Format