RIGOL
2-114
DS4000E Programming Guide
:MASK:Y
Syntax
:MASK:Y <y>
:MASK:Y?
Description
Sets or queries the vertical tolerance range of the pass/fail test mask. The default unit is div (in the vertical
direction).
Parameter
Name
Type
Range
Default
<y>
Real
0.03125 to 4 (with a step of 0.03125)
0.375
Remarks
For the vertical tolerance range, the oscilloscope only displays the number to two decimal places
(round-up). For example, when the vertical tolerance range is 0.0625 div, 0.06 is displayed; when the
vertical tolerance range is 0.125 div, 0.13 is displayed.
Return Format
The query returns the vertical tolerance range in scientific notation.