![Rigol DS4000E Series Programming Manual Download Page 156](http://html1.mh-extra.com/html/rigol/ds4000e-series/ds4000e-series_programming-manual_1460996156.webp)
RIGOL
2-144
DS4000E Programming Guide
:MEASure:RDELay:SAVerage
:MEASure:RDELay:SCURrent
:MEASure:RDELay:SDEViation
:MEASure:RDELay:SMAXimum
:MEASure:RDELay:SMINimum
Syntax 1
:MEASure:RDELay:SAVerage [<chanA>,<chanB>]
:MEASure:RDELay:SCURrent [<chanA>,<chanB>]
:MEASure:RDELay:SDEViation [<chanA>,<chanB>]
:MEASure:RDELay:SMAXimum [<chanA>,<chanB>]
:MEASure:RDELay:SMINimum [<chanA>,<chanB>]
Description 1
Measures the delay (rising edge-rising edge) between the two specified source channels, and enables the
statistic function of the measurement results.
Syntax 2
:MEASure:RDELay:SAVerage? [<chanA>,<chanB>]
:MEASure:RDELay:SCURrent? [<chanA>,<chanB>]
:MEASure:RDELay:SDEViation? [<chanA>,<chanB>]
:MEASure:RDELay:SMAXimum? [<chanA>,<chanB>]
:MEASure:RDELay:SMINimum? [<chanA>,<chanB>]
Description 2
Queries the average value, current value, standard deviation, maximum value, or minimum value of the
delay (rising edge-rising edge) measurement results when the delay (rising edge-rising edge)
measurement function between the two specified source channels and the statistic function of the
measurement results are enabled. The default unit is s.
Parameter
Name
Type
Range
Default
<chanA>
Discrete
{CHANnel1|CHANnel2|CHANnel3|CHANnel4}
--
<chanB>
Discrete
{CHANnel1|CHANnel2|CHANnel3|CHANnel4}
--
Remarks
Before making measurements, turn on the channels to be measured.
When [<chanA>,<chanB>] is omitted, this command will be applied to Source A (refer to
the :MEASure:SETup:DSA command) and Source B (refer to the :MEASure:SETup:DSB command) of
delay measurement currently set by default.
Return Format
The query returns the average value, current value, standard deviation, maximum value, or minimum value
in scientific notation. When the measurement results are invalid, the query returns 9.9e37.