:MASK:SOURce
Syntax
:MASK:SOURce <source>
:MASK:SOURce?
Description
Sets or queries the source channel of the pass/fail test.
Parameter
Name
Type
Range
Default
<source>
Discrete
{CHANnel1|CHANnel2|CHANnel3|CHANnel4}
CHANnel1
Remarks
Only the currently enabled channel (refer to the :CHANnel<n>:DISPlay command) can be selected as the
source channel.
Return Format
The query returns CHAN1, CHAN2, CHAN3, or CHAN4.
:MASK:TOTal?
Syntax
:MASK:TOTal?
Description
Queries the total number of frames in the pass/fail test results.
Return Format
The query returns an integer.
:MASK:X
Syntax
:MASK:X <x>
:MASK:X?
Description
Sets or queries the horizontal tolerance range of the pass/fail test mask. The default unit is div (in the
horizontal direction).
Parameter
Name
Type
Range
Default
<x>
Real
0.02 to 4 (with a step of 0.02)
0.24
Return Format
The query returns the horizontal tolerance range in scientific notation.