Troubleshooting
64
Problem Possible
Cause
Instructions
Alarm Sound
Continuous, No Lights
No power, battery drained.
Connect power cord/battery
charger. Use Power-Off to silence
alarm.
Press and hold the blue
RESET
button.
Press and release the gray
POWER
button. Wait 2
seconds, then release the
RESET
button.
Prior to use, allow battery to charge
approximately 30 minutes. You may
then operate the monitor while it is
plugged in. Allow the battery to
charge for 6 hours before using the
monitor on battery power.
Internal part failure.
Contact Respironics.
Alarm sounds weak.
Internal part failure.
Low battery
Contact Respironics.
Charge battery.
Loose lead
Light remains on.
Connections between
electrodes, lead wires, and
patient cable are not properly
made.
Defective lead wires.
Defective patient cable.
Defective electrodes.
Internal part failure.
Verify that
(a) baby’s skin underneath
electrodes is clean,
(b) electrodes are clean, and
(c) lead wires are fully inserted into
the electrodes and patient
cable.
Replace lead wires and perform
Functional Self-Test.
Replace patient cable and perform
Functional Self-Test.
Replace electrodes.
Contact Respironics.
Unable to download via
modem.
AC power not connected to
SmartMontior 2.
Plug the SmartMonitor 2 into AC
power before performing a modem
download.
Unable to communicate
by direct connect
Download cable was not
connected to the SmartMonitor
2 and the computer prior to
turning on the SmartMonitor 2.
Connect the download cable to the
SmartMonitor 2 and the computer
before turning the SmartMonitor 2
on.
Summary of Contents for SmartMonitor2
Page 1: ...Professional Operator s Manual ...
Page 6: ...Table of Contents ...
Page 12: ...Introduction 6 ...
Page 16: ...Warnings and Cautions 10 ...
Page 36: ...Using Auxiliary Equipment 30 ...
Page 38: ...Dealer Preparation for Home Setup 32 ...
Page 42: ...Home Setup 36 Baby s right side Baby s left side ...
Page 50: ...Reducing False Alarms 44 ...
Page 54: ...Monitoring the Baby 48 ...
Page 58: ...Transferring SmartMonitor 2 Information 52 ...
Page 64: ...Transferring SmartMonitor 2 Information 58 ...
Page 68: ...Caring for the SmartMonitor 2 62 ...
Page 72: ...Ordering Information 66 ...
Page 74: ...Appendices 68 ...
Page 76: ...Appendices 70 ...
Page 78: ...Specifications 72 ...
Page 80: ...H577 4000 10 KK 5 3 06 ...