Portable Operation of the SmartMonitor 2
49
Portable Operation of the SmartMonitor 2
The SmartMonitor 2 is designed for portable use. When the power cord is not used, the monitor relies on
a previously charged internal battery for power.
Respironics recommends that the SmartMonitor 2 be used with the power cord/battery charger whenever
possible. However, when the monitor is used without the power cord/battery charger the SmartMonitor 2
is fully functional. All alarms are operational. With a fully charged battery, the monitor will run for 15
hours. The amount of time to completely recharge a fully depleted battery is 6 hours.
Charging the SmartMonitor 2 Battery
As a rule, a fully charged battery can operate for 15 hours. This may vary, however, depending on the
level of use, number of alarms, and other factors. When the low battery light comes on, you should
recharge the battery immediately. A fully drained battery should be recharged for 6 hours. When you
need to recharge the SmartMonitor 2 battery, follow the steps below.
Connect the power cord/battery charger to the back panel of the SmartMonitor 2.
Plug the power cord/battery charger into a power outlet. The green
charger
light comes on.
If the SmartMonitor 2 is turned on, the yellow
low battery
light blinks until the minimum
charge level is reached. Then, the yellow light goes off.
Fully drained batteries need about 6 hours to recharge
.
Unit should be at room temperature 65
°
F to 85
°
F while charging or batteries may not
reach a full charge. Only use Respironics batteries.
The Power Cord Connector must be plugged into the SmartMonitor 2 DC Power
Input as shown in the illustration above. The Power Cord Connector can only be
inserted as shown above.
Summary of Contents for SmartMonitor2
Page 1: ...Professional Operator s Manual ...
Page 6: ...Table of Contents ...
Page 12: ...Introduction 6 ...
Page 16: ...Warnings and Cautions 10 ...
Page 36: ...Using Auxiliary Equipment 30 ...
Page 38: ...Dealer Preparation for Home Setup 32 ...
Page 42: ...Home Setup 36 Baby s right side Baby s left side ...
Page 50: ...Reducing False Alarms 44 ...
Page 54: ...Monitoring the Baby 48 ...
Page 58: ...Transferring SmartMonitor 2 Information 52 ...
Page 64: ...Transferring SmartMonitor 2 Information 58 ...
Page 68: ...Caring for the SmartMonitor 2 62 ...
Page 72: ...Ordering Information 66 ...
Page 74: ...Appendices 68 ...
Page 76: ...Appendices 70 ...
Page 78: ...Specifications 72 ...
Page 80: ...H577 4000 10 KK 5 3 06 ...