Transferring SmartMonitor 2 Information
54
Modem Auto Dial
The SmartMonitor 2 must be connected to AC power during modem download.
To have SmartMonitor 2 call the computer for a download it must be programmed with a Host Phone
Number, which is the phone number to be called for a connection with the computer as well as the date
and time to call. If the phone number field is blank, no attempt to call will be made. These must be set up
in the SmartMonitor 2 before it is placed in the patient home. When the preset time for download
approaches or when the memory full light illuminates, follow these instructions:
If there is an emergency and access to the telephone is required while the modem is in use,
remove phone cord from modem slot and use a working phone.
Plug the telephone wire into the modem connector on the
back of the SmartMonitor 2 (see illustration).
Plug the other end of the phone wire into the wall phone jack.
Refer to the Synergy-E software for instruction on receiving
a call from the SmartMonitor 2.
Modem will continue calling every 30 minutes, for 4 hours
until it connects with the computer. If no connection is made,
the modem will try again the next day at the pre-selected
time. For example, if the modem is programmed to call at
7am but fails to make a connection after 8 attempts it will
stop trying. The next day at 7am it will try again. This will
continue until a connection is made.
Modem Auto Answer in Communications Mode
The SmartMonitor 2 must be connected to AC power during modem download.
1. Connect a phone cord from the wall jack to the modem connector on the back of the SmartMonitor 2.
2. To allow the SmartMonitor 2 to work with the modem, place the monitor in Communications Mode:
Press and release the
POWER
button to turn the SmartMonitor 2 on.
Press and hold the blue
RESET
button.
Wait until the monitor alarms.
Release
the
RESET
button. Briefly press and release the
RESET
button again. The SmartMonitor
2 display will read “Monitor is in Communication Mode”.
The SmartMonitor 2 beeps every 10 seconds whenever it is in the Communications Mode. This is a
reminder that the SmartMonitor 2 is powered on for working with the computer, memory card, or modem,
but
not
for monitoring the infant.
Do not connect the infant to the SmartMonitor 2 when in the Communications Mode, the apnea
and heart rate alarms are not operational in this mode.
The dealer should now call to retrieve the information. The phone may ring, but do not answer.
The modem answers the call and connects the SmartMonitor 2 to the computer through the
phone line.
4. The amount of time to transfer the information will vary. When the transfer is complete the
Summary of Contents for SmartMonitor2
Page 1: ...Professional Operator s Manual ...
Page 6: ...Table of Contents ...
Page 12: ...Introduction 6 ...
Page 16: ...Warnings and Cautions 10 ...
Page 36: ...Using Auxiliary Equipment 30 ...
Page 38: ...Dealer Preparation for Home Setup 32 ...
Page 42: ...Home Setup 36 Baby s right side Baby s left side ...
Page 50: ...Reducing False Alarms 44 ...
Page 54: ...Monitoring the Baby 48 ...
Page 58: ...Transferring SmartMonitor 2 Information 52 ...
Page 64: ...Transferring SmartMonitor 2 Information 58 ...
Page 68: ...Caring for the SmartMonitor 2 62 ...
Page 72: ...Ordering Information 66 ...
Page 74: ...Appendices 68 ...
Page 76: ...Appendices 70 ...
Page 78: ...Specifications 72 ...
Page 80: ...H577 4000 10 KK 5 3 06 ...