Home Setup
35
Step 5: Attach the Electrodes to the Infant Belt.
Place the electrode belt on a flat surface.
Lay the baby on the belt so the belt is aligned with
the baby’s nipples (see illustration).
Place the electrodes, Velcro-side down, on either
side of the belt as follows:
Place the electrode with the
white
lead wire on
the
baby’s
right
side.
Place the electrode with the
black
lead wire on
the
baby’s
left
side.
Place the electrodes far enough apart so that when
the belt is wrapped around the baby, the electrode
will be located along the mid-line of the side just
below or lined up with the nipples.
Be sure the lead wires and patient cable are leading
down and away from the baby's face and neck (see
illustration).
Baby’s
right side
Baby’s
left side
The white lead wire location
is illustrated with a white
electrode.
Step 6: Wrap the Electrode Belt around the Baby.
Wrap the belt around the baby’s
chest and fasten it with the white
Velcro tab.
The belt should be snug enough
so that you can only insert two of
your fingers (with your hand lying
flat against baby) between the belt
and the baby.
With newborns and very small
babies, you may need to shorten the
belt by cutting off a part of it. Be sure
to leave enough room to fasten the belt
securely.
Summary of Contents for SmartMonitor2
Page 1: ...Professional Operator s Manual ...
Page 6: ...Table of Contents ...
Page 12: ...Introduction 6 ...
Page 16: ...Warnings and Cautions 10 ...
Page 36: ...Using Auxiliary Equipment 30 ...
Page 38: ...Dealer Preparation for Home Setup 32 ...
Page 42: ...Home Setup 36 Baby s right side Baby s left side ...
Page 50: ...Reducing False Alarms 44 ...
Page 54: ...Monitoring the Baby 48 ...
Page 58: ...Transferring SmartMonitor 2 Information 52 ...
Page 64: ...Transferring SmartMonitor 2 Information 58 ...
Page 68: ...Caring for the SmartMonitor 2 62 ...
Page 72: ...Ordering Information 66 ...
Page 74: ...Appendices 68 ...
Page 76: ...Appendices 70 ...
Page 78: ...Specifications 72 ...
Page 80: ...H577 4000 10 KK 5 3 06 ...