Home Setup
34
Step 3: Connect the Lead Wires to the Patient Cable.
The larger end of the patient cable has
three openings, marked
LA
(black),
RL
(green), and
RA
(white).
Take the
white
lead wire and
insert it into the opening marked
RA
.
Take the
black
lead wire and
insert it into the opening marked
LA
.
Firmly push each lead wire in until
the socket snaps into place.
When you need to remove a
lead wire, grasp and pull at the
strain relief area located near
the connecting tip.
Do not
grasp the wire.
Use of the third (green - RL) electrode and
lead wire is normally not required, but may
help reduce excessive false low heart rate
alarms.
Step 4: Connect the Lead Wires to the Electrodes.
Insert the black
LA
lead wire into one
electrode.
Insert the white
RA
lead wire into the
other electrode.
Make sure the metal
tips of the lead wires
are fully inserted into
the electrodes.
Strain Relief
Area.
Summary of Contents for SmartMonitor2
Page 1: ...Professional Operator s Manual ...
Page 6: ...Table of Contents ...
Page 12: ...Introduction 6 ...
Page 16: ...Warnings and Cautions 10 ...
Page 36: ...Using Auxiliary Equipment 30 ...
Page 38: ...Dealer Preparation for Home Setup 32 ...
Page 42: ...Home Setup 36 Baby s right side Baby s left side ...
Page 50: ...Reducing False Alarms 44 ...
Page 54: ...Monitoring the Baby 48 ...
Page 58: ...Transferring SmartMonitor 2 Information 52 ...
Page 64: ...Transferring SmartMonitor 2 Information 58 ...
Page 68: ...Caring for the SmartMonitor 2 62 ...
Page 72: ...Ordering Information 66 ...
Page 74: ...Appendices 68 ...
Page 76: ...Appendices 70 ...
Page 78: ...Specifications 72 ...
Page 80: ...H577 4000 10 KK 5 3 06 ...