Caring for the SmartMonitor 2
61
(heart) light remains on. (There should be no green
heart
or
respiration
light flashes during this
time).
Follow the instructions in the
Self-Test Troubleshooting
section, if necessary.
Remove the lead wires from the
functional self-test
socket. To remove lead wires, grasp and pull
the strain relief area located near the connecting tip.
Do not grasp the wire.
The
loose lead
light will come on, and the alarm changes from beeping to continuous. This lets you
know the SmartMonitor 2, patient cable, and lead wires are working properly.
Now turn the monitor off.
•
Press and hold the blue
RESET
button.
•
Press and release the gray
POWER
button.
•
Wait 2 seconds, and then release the
RESET
button.
Self-Test Troubleshooting
Follow the instructions given below if any of the conditions described occur. Start the test over once the
problem has been corrected.
Alarm
Condition Solution
Low Battery
If the
low battery
light stays on
longer than half a minute, the
batteries are completely
discharged.
Turn the monitor off using the
correct Power Off procedure
described in the Turning the Monitor
Off – Sibling Alarm section in this
manual.
Make sure the power cord/battery
charger is plugged into a live power
outlet and is properly connected to
the monitor. (See
Charging the
Battery
for more information).
Allow the battery to charge
approximately 30 minutes. You may
then operate the monitor while it is
plugged in. Allow the battery to
charge for 6 hours before using the
monitor on battery power.
Memory Full
The SmartMonitor 2 memory
has violated the Memory Full
parameter.
Press the
RESET
button to silence the
alarm. The SmartMonitor 2 memory
needs to be transferred Self-Test may
continue.
Loose Lead
Indicates loose or bad
electrodes, lead wires, and
patient cable.
Check all connections and/or replace
lead wires first, then the patient cable if
necessary.
The SmartMonitor 2 lights and alarms should respond as described above. If not, contact
Respironics Service before using the unit to monitor a baby.
Do not use the monitor if the alarm sounds weak or does not activate upon initial power-on.
Summary of Contents for SmartMonitor2
Page 1: ...Professional Operator s Manual ...
Page 6: ...Table of Contents ...
Page 12: ...Introduction 6 ...
Page 16: ...Warnings and Cautions 10 ...
Page 36: ...Using Auxiliary Equipment 30 ...
Page 38: ...Dealer Preparation for Home Setup 32 ...
Page 42: ...Home Setup 36 Baby s right side Baby s left side ...
Page 50: ...Reducing False Alarms 44 ...
Page 54: ...Monitoring the Baby 48 ...
Page 58: ...Transferring SmartMonitor 2 Information 52 ...
Page 64: ...Transferring SmartMonitor 2 Information 58 ...
Page 68: ...Caring for the SmartMonitor 2 62 ...
Page 72: ...Ordering Information 66 ...
Page 74: ...Appendices 68 ...
Page 76: ...Appendices 70 ...
Page 78: ...Specifications 72 ...
Page 80: ...H577 4000 10 KK 5 3 06 ...