Responding to Alarms
41
Responding to System Alarms
A System Alarm indicates that the monitor may not be functioning properly or at optimum capacity. The
information in this section will help you respond appropriately to system alarms. When a SmartMonitor 2
system alarm occurs, one of the lights at the bottom of the front panel will come on.
If this light is
on
And this condition
exists
It means...
Power
Continuous green
light, no alarm.
Normal Operation. The green
power
indicator light will come
on and stay on for as long as the monitor is on.
Charger
Continuous green
light, no alarm.
Normal Operation. The green
charger
light will come on and
stay on while the battery charger is plugged into an active
outlet and connected to the SmartMonitor 2.
Low Battery
Flashing yellow light,
continuous alarm.
Continuous yellow
light, continuous
alarm.
This is a warning that the battery voltage is very low and
should be recharged soon. (
See
Charging the Battery
in this
manual). Press the
RESET
button to temporarily silence the
alarm. The alarm will resound in 2 minutes if the monitor has
not been plugged in. The yellow light will continue to flash.
This is a warning that the battery is too low for the
SmartMonitor 2 to operate properly. The SmartMonitor 2 must
be recharged. Turn the monitor off. Then, recharge the
battery. (
See
Charging the Battery
in this manual).
If you do not recharge the battery, the system will
automatically shut down.
Memory Full
Flashing yellow light,
continuous alarm
When the SmartMonitor 2’s Memory Almost full parameter is
reached, the Memory Full light will flash. The alarm will sound
continuously.
The alarm will sound only if the dealer programs
the SmartMonitor 2 to do so at the 50% full or at 80% full.
Press the
RESET
button to silence the alarm. The light will
blink every second.
Memory
full
is a warning condition. You can continue
monitoring
.
The caregiver should contact the dealer to
resolve the alarm.
Memory Full
Continuous yellow
light, continuous alarm
SmartMonitor 2 memory is 100% full. Press the
RESET
button
to silence the alarm.
The alarm will sound only if the dealer programs the
SmartMonitor 2 to do so.
The light will stay on continuously. The caregiver should
contact the dealer to resolve the alarm.
(
See
Transferring the
Monitor Memory
in this manual for more
information).
Summary of Contents for SmartMonitor2
Page 1: ...Professional Operator s Manual ...
Page 6: ...Table of Contents ...
Page 12: ...Introduction 6 ...
Page 16: ...Warnings and Cautions 10 ...
Page 36: ...Using Auxiliary Equipment 30 ...
Page 38: ...Dealer Preparation for Home Setup 32 ...
Page 42: ...Home Setup 36 Baby s right side Baby s left side ...
Page 50: ...Reducing False Alarms 44 ...
Page 54: ...Monitoring the Baby 48 ...
Page 58: ...Transferring SmartMonitor 2 Information 52 ...
Page 64: ...Transferring SmartMonitor 2 Information 58 ...
Page 68: ...Caring for the SmartMonitor 2 62 ...
Page 72: ...Ordering Information 66 ...
Page 74: ...Appendices 68 ...
Page 76: ...Appendices 70 ...
Page 78: ...Specifications 72 ...
Page 80: ...H577 4000 10 KK 5 3 06 ...